4.4
TERMINAL ASSIGNMENT
term. blk.
803
804
804
804
806
812
813
*
mentioned state of relay valid only for switched-off instrument.
** when wiring EMC-ompatible acc. to CE requirements, please connect the
screen of the loadcell extension cable to the separate earthing clamp of
bold B (see also fig. 7.8).
4.5
MAINS CONNECTION
The instrument is set up at the factory for 220 V AC-operation. For change over
to other mains voltages proceed in accordance with para. 5.1.
Please note that the instrument has no mains switch, and is therefore ready for
operation as soon as the mains supply is connected to it.
According to EN 61010-1, part 6.12.2.1, it must be possible to switch off the
mains voltage by means of a appropriate device situated close to the instrument.
This device could be e.g. a mains connector, correspondingly marked, and
easily to be reached by the operator.
terminal
max. diameter
1
}
screen
2
}
see
}
3
}
PR 6135
4
}
loadcell
5
}
extension
6
}
cable
7
}
8
}
solid
# 4 mm²,
9
}
10
}
stranded
# 2.5 mm²
11
}
solid # 4 mm²,
1
}
stranded # 2.5 mm²
2
}
1
}
solid
# 4 mm²,
2
}
stranded # 2.5 mm²
3
}
1
}
solid
# 4 mm²
2
}
stranded # 2.5 mm²
3
}
Function
(not to be used!) **
+ measuring signal
- measuring signal
+ 6 V Loadcell supply
+ 6 V Sense voltage
- 6 V Sense voltage
- 6 V Loadcell supply
+ } 0 ..+10 V
- } output voltage.
- } 0 ..+20, or 4 ..+20 mA
+ } output current.
neutral
live
change over
make contact * } < LIM 2 (MIN)
break contact * }
change over
make contact * } > LIM 1 (MAX)
break contact * }
Page E 4-2
} mains
} connection
}
}