T E C H N I C A L D A T A
3.1
PERFORMANCE CHARACTERISTICS / SPECIFICATIONS
Properties expressed in numerical values with stated tolerance are guaranteed
by the manufacturer. Specified non-tolerance numerical values indicate those
that could be nominally expected from the mean of a range of identical
instruments. This specification is valid after the instrument has warmed up for
2 hours (reference temperature +23 E C).
If not stated otherwise, relative or absolute tolerances relate to the set value.
3.2
GENERAL DATA
Typ
Order Number
Measuring principle
Supply voltage
Connection
Voltage
Tolerance
Power consumption
Frequency
Safety class
3.3
SPECIFICATION
3.3.1 Loadcell connection
Type of loadcell
Loadcell connection
Cable length
PR 1591/00, PR 1591/10
9405 115 91001, 9405 115 91101
DC-amplification, based on loadcell supply
symmetrically to earth and differential sense
amplifier.
Live and Neutral via 2-pole screw terminal,
protective earth via separate earthing bold.
Terminals suitable for 2.5 mm².
110, 128, 220 or 238 V AC (nominal);
settable via solder points close to the
mains transformer
-15 %/+10 %
8 W at max. (12 VA)
50/60 Hz
I (protective lead) acc. to IEC 536.
all types of strain-gauge-loadcells or any
other sources with linear voltage output.
(sensitivity range 0.2mV/V ... 2mV/V).
via screw terminals, suitable for 2.5 mm² at
max., to be connected in 6-wire technique for
standard installations.
Alternatively also in 4-wire technique
possible.
300 m at max.