Loadcell supply voltage
Tolerance
Loading
Sense input
Input voltage
Max. input voltage
Max. common mode voltage
Input resistance
Measuring signal input
SPAN coarse
SPAN fine
Accuracy
Ri
Max. input voltage
Common mode rejection
Dead load suppression
ZERO-coarse
ZERO-fine
Accuracy
Filtering of measuring
voltage
Limit frequency
3.3.2 Outputs
Voltage output
Nominal range
Max. load
Connection
Cable length
12 V DC, symmetrically to earth,
short circuit proof.
±5 %
$ 100 Ohm (= 6 loadcells of 600 Ohm each,
connected in parallel).
differential input, sensing the loadcell supply
voltage at the measuring point (junction box).
±6 V (12 V DC diff.)
15 V
15 V
34 kOhm
0.2 mV/V ... 2.0 mV/V nominal at 12 V.
within 2.5 mV ... 25 mV nominal by means of
an incorporated 8-stage DIL-switch.
approx. ±5 % of f.s.d., by means of an
incorporated multi-turn potentiometer.
<0.05 %
>10 Mohm
15 V
>110 dB
0... 15 mV nominal, settable by means of an
incorporated 7-stage DIL switch.
0 ... 0.19 mV nominal at Ue= 2.5 mV
( = 0.208 mV/V).
< 1 µV
No interaction between SPAN and ZERO.
by means of an active 2-stage filter, to
be switched on by closing a solder-link.
f without filter = 630 Hz, f with filter = 1 Hz.
g
0... 10 V DC / 5 mA at max.
$ 2 kOhm
via screw terminals of 2.5 mm² at max.
50 meter at max.