B. Braun Aesculap Spine S4 Cervical FW086SU Gebrauchsanweisung/Technische Beschreibung Seite 19

Navigierte instrumente
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 26
Validated procedure
Specific requirements
Manual pre-cleaning
with ultrasound and
brush, and subsequent
mechanical alkaline
cleaning and thermal
disinfection
FJ985R
FW661R
Manual cleaning/disinfection
Prior to manual disinfecting, allow water to drip off for a sufficient
length of time to prevent dilution of the disinfecting solution.
After manual cleaning/disinfection, check visible surfaces visually for
residues.
Repeat the cleaning /disinfection process if necessary.
Manual cleaning with immersion disinfection
Phase
Step
I
Disinfecting cleaning
II
Intermediate rinse
III
Disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning brushes and disposable
syringes, see Validated cleaning and disinfection procedure.
Phase I
Fully immerse the product in the cleaning/disinfectant for at least 15
min. Ensure that all accessible surfaces are moistened.
Clean the product with a suitable cleaning brush in the solution until
all discernible residues have been removed from the surface.
If applicable, brush through non-visible surfaces with an appropriate
cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during
cleaning.
Cleaning brush: 30 mm/∅: 4.5 mm, e.g. TA011944 and cleaning
brush: 20 mm/∅: 2.5 mm, e.g. TE654202 and cleaning brush:
50 mm/∅: 10 mm, e.g. TA007747
20 ml disposable syringe
Place the product in a tray that is suitable for cleaning (avoiding rins-
ing blind spots).
Keep working ends open for cleaning.
Place products in the tray with their hinges open.
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold) >15
2
RT (cold) 1
-
RT (cold) 15
2
RT (cold) 1
-
RT
-
-
Water
Chemical
quality
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
D–W
-
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
FD-W
-
-
-
Thoroughly rinse through these components with the cleaning disin-
fectant solution (at least five times), using a disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under run-
ning water.
Mobilize non-rigid components, such as set screws, joints, etc. during
rinsing.
Drain any remaining water fully.
Reference
Chapter Mechanical cleaning/disinfection with
manual pre-cleaning and sub-chapter:
Chapter Manual pre-cleaning with ultra-
sound and brush
Chapter Mechanical alkaline cleaning and
thermal disinfecting
17