3.11 Mechanical cleaning/disinfection with manual pre-cleaning

Note
The cleaning and disinfection device must be of tested and approved effec-
tiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced
and checked at regular intervals.

3.11.1 Manual pre-cleaning with a brush

Phase
Step
I
Disinfecting cleaning
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed fresh
Note the information on appropriate cleaning brushes and disposable
syringes, see Validated cleaning and disinfection procedure.
Phase I
Fully immerse the product in the cleaning/disinfectant for at least 15
min. Ensure that all accessible surfaces are moistened.
Clean the product with a suitable cleaning brush in the solution until
all discernible residues have been removed from the surface.
If applicable, brush through non-visible surfaces with an appropriate
cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during
cleaning.
Thoroughly rinse through these components with the cleaning disin-
fectant solution (at least five times), using a disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under run-
ning water.
Mobilize non-rigid components, such as set screws, joints, etc. during
rinsing.
16
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
>15
2
RT (cold)
1
-
Water
Chemical
quality
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
D–W
-