Aesculap
Reciprocating saw GA334
Phase VI
Rinse disinfected surfaces after the prescribed reaction time for at
least 1 minute under running demineralized water.
Drain any remaining water fully.
7.7

Automatic cleaning/disinfection with manual pre-cleaning

Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
Note
The cleaning and disinfection machine used for processing must be ser-
viced and checked at regular intervals.
7.7.1

Manual pre-cleaning with a brush

Phase
Step
I
Rinsing
II
Brushes
D–W:
Drinking water
RT:
Room temperature
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Push the saw blade coupling 14 towards the housing as far as it will
go.
Phase I
Mobilize non-rigid components during cleaning, e.g. pull locking
sleeve 20 back from the support foot 16.
Thoroughly clean the product under running water.
Use cleaning template 15 to clean the slot of the saw blade
coupling 14. Insert cleaning template into the slot and carefully push
out the debris through the slot or the inspection holes at least 5 times
in both directions. Turn template 180° in order to do so.
Manually clean the contact sites of the rinsing adapter GA674890 and
the rotating sleeve 12 thoroughly.
10
Acculan 4
®
T
t
[°C/°F]
[min]
RT (cold)
-
RT (cold)
-
Stage VII
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).
Conc.
Water
Chemistry/Note
[%]
quality
-
D–W
Until visually clean
-
D–W
Until visually clean
Phase II
Mobilize non-rigid components during cleaning, e.g. pull locking
sleeve 20 back from the support foot 16.
Brush difficult to access surfaces with a suitable plastic cleaning brush
for at least 1 min.
After manual preparation, check visible surfaces for residue and repeat
the pre-cleaning process as needed.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).