Troubleshooting

Problem
The signal lights do
not turn on.
The battery light
turns off.
The programme
display lights are on,
but not blinking.
The programme dis-
play lights work, but
no pulses are sent for
muscle stimulation.
00069_DE_GB_FR_NL_V2.indd 31

Possible Cause

No batteries have been
inserted.
The batteries have not been
inserted properly.
The batteries are too weak.
The silicone electrodes are not
in contact with your skin.
The microprocessor unit has
not been properly secured to
the belt.
The intensity has been set to
the lowest level.
The microprocessor unit is not
working properly.
The intensity level is too low.
The belt has not been put on
properly.
The silicone electrode or
stimulation pad are worn
down or dirty.
The skin is too dry
31
Solution
Insert batteries and press
ON/HI.
When inserting the batteries,
ensure that the polarity is
correct.
Insert new batteries.
Make sure that the stimulation
pads are positioned properly
and the belt is somewhat
tighter.
Check whether the
microprocessor unit has been
securely attached to the snap
fasteners.
Push
ON/HI
in order to
increase the intensity.
Turn the device off, take out
the battery for 1 minute and
attempt to restart the device.
Push the
ON/HI
button
several times to increase the
intensity.
Shift the belt until you have
found the position with the
strongest muscle contraction.
Replace or clean the silicone
electrode or stimulation pad.
Apply more gel on the skin.
16.12.09 12:33