10. Technical Service

DANGER
Danger to life of patients and users in case of malfunctions and/or
failure of protective measures!
Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.
CAUTION
Modifications of medical devices may result in a loss of potential
guarantee/warranty claims and forfeiture of applicable licenses.
Do not modify the product.
Contact national B. Braun/Aesculap representative for service and
repair.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the address indicated above.

11. Accessories/Spare parts

Art. no.
Name
GA346
Li-Ion battery short
GA340345
Nut
GA643817
Flap rod
GA670210
Flap
GA675
Lid
GA678
Sterile funnel
GA679
Battery removal device
GB228R
Dermatome blade
GB256R
ECCOS set for dermatome
GB487R
ECCOS fixation for lid
GB488R
ECCOS fixation for sterile funnel
GB489R
ECCOS holder for battery removal device
GB498R
ECCOS fixation for straight machine
GB600
STERILIT Power Systems oil spray
TA014548
Instructions for use of Dermatome GA340/GA341
(A4 for ring binder)
TA014549
Instructions for use of Dermatome GA340/GA341
(flyer)
12

12. Technical data

12.1 Classification according to Regulation (EU)
2017/745
Art. no.
Name
GA340
Dermatome
GA341
0.1 mm Dermatome

12.2 Performance data, information about standards

Max. power
Max. oscillation frequency
Blade travel
Cutting depth setting
Cutting width
Weight (ready for operation)
Dimensions (L x W x H, ready for
operation)
Application part
EMC
Compliance with standards
The product was subject to 500 preparation cycles as a test by the manu-
facturer and has passed.

12.3 Nominal operating mode

Operation with non-periodic load and speed changes (type S9 pursuant to
IEC EN 60034-1)
60 second application, 60 second pause
10 repetitions
30 min cooling time
Max. Temperature 48 °C

12.4 Ambient conditions

Operation
Temperature
10 °C to 27 °C
Relative humid-
30 % to 75 %
ity
Atmospheric
700 hPa to 1 060 hPa
pressure
Class
IIa
IIa
approx. 250 W
6 500 rpm
approx. 3.1 mm
0.2 mm to 1.2 mm (with GA340)
0.1 mm to 1.1 mm (with GA341)
Scale gradations 1/10 mm
max. approx. 78 mm, set by
10 flaps of width 8 mm
1.65 kg ±10 %
277 mm x 113 mm x 83 mm ±5 %
Type BF
IEC/DIN EN 60601-1-2
IEC/EN 60601-1
Transport and storage
-10 °C to 50 °C
10 % to 90 %
500 hPa to 1 060 hPa