Issue
Cause
Sterile funnel or battery
Sterile funnel or battery
removal aid cannot be
removal device not compati-
plugged in
ble
Sterile funnel or battery
removal device
deformed/defective
Battery compartment on
product deformed/defective
Speed control knob can-
Product in locked OFF position Safety catch is in OFF position
not be actuated
Speed control knob
jammed/defective
Saw blade not moving
Loud running noise from
motor
Insufficient cutting per-
Saw blade teeth worn
formance of the saw
Battery too weak
blade
Battery worn/defective
Saw blade overheated
The saw blade cannot be
Coupling components or saw
coupled
blades deformed
-
Saw blade cannot be
Locking pin not fully releasing
uncoupled
the saw blade

10. Technical Service

DANGER
Danger to life of patients and users in case of malfunctions and/or
failure of protective measures!
Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.
CAUTION
Modifications of medical devices may result in a loss of potential
guarantee/warranty claims and forfeiture of applicable licenses.
Do not modify the product.
Contact national B. Braun/Aesculap representative for service and
repair.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the address indicated above.
12
Detection
Sterile funnel or battery removal aid
not plugged into battery compartment
Sterile funnel or battery removal device
difficult or not attachable
Difficult or impossible to plug in the
sterile funnel or battery removal aid
Speed control knob not working
Broken gear unit
Blunt saw blade
Insufficient motor power/speed
Chips not removed adequately
Deformation
Incompatible saw blade
Push button for unlocking saw blade
not fully pushed

11. Accessories/Spare parts

Art. no.
GA675
GA676
GA678
GA679
GB243R
GB244R
GB600
GB495R
GB487R
GB488R
GB489R
TA014542
TA014543
Remedy
Use suitable sterile funnel or battery removal aid
for GA331.
Replace sterile funnel or battery removal aid.
Have the product repaired by the manufacturer.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Have repaired by manufacturer
Replace saw blade
Charge battery
Replace battery
When sawing, move back and forth more and rinse
sufficiently
Have repaired by manufacturer
Insert a compatible Aesculap saw blade, see
Accessories/Spare parts
Fully press the saw blade release button, see
Attaching and detaching the saw blade
Name
Lid
NiMH battery long
Sterile funnel
Battery removal aid
ECCOS screen basket with holder for Acculan 4
STERILIT Power Systems oil spray
ECCOS fixation for machine
ECCOS holder for lid
ECCOS holder for sterile funnel
ECCOS holder for battery removal aid
Instructions for use for oscillating saw GA331 (A4
for ring binder)
Instructions for use for oscillating saw GA331
(flyer)