B. Braun Aesculap Acculan 4 GA340 Gebrauchsanweisung/Technische Beschreibung Seite 13

Dermatom
Vorschau ausblenden Andere Handbücher für Aesculap Acculan 4 GA340:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 16
Issue
Cause
Insufficient power
Product defective
Blunt tool
Loud running noise
Defective product drive/ball
bearings
Lid cannot be fit-
Lid not compatible
ted/removed
Lid deformed/broken
Connection to product
deformed/defective
Battery cannot be
Battery not compatible
inserted/removed
Battery deformed/defective
Battery compartment on
product deformed/broken
Sterile funnel or battery
Sterile funnel or battery
removal device cannot be
removal device not compati-
attached
ble
Sterile funnel or battery
removal device deformed/bro-
ken
Battery shaft to product
deformed/defective
Speed control knob not
Product in locked OFF position Safety catch is in OFF position
working
Speed control knob sticks/is
defective
Dermatome blade not
Defective gearings
moving
Insufficient cutting per-
Dermatome blade blunted
formance of dermatome
Dermatome maladjusted
blade
Cutting depth
Skin grafts uneven/too
uneven/deviating from
thick/too thin
setting
Detection
Insufficient product performance
Steep rise in temperature after a short
period of time
Tool blades worn out
Loud, perceptible noise during opera-
tion
Lid does not click into place
Lid is hard to fit/remove or cannot be
fitted/removed
Lid is difficult or cannot be
mounted/removed
Battery cannot be completely inserted
into the battery shaft
Battery is hard to fit/remove or cannot
be fitted/removed
Battery is difficult or cannot be
mounted/removed
Sterile funnel or battery removal device
cannot be attached to the battery com-
partment
Sterile funnel or battery removal device
difficult to attach or cannot be
attached
Sterile funnel or battery removal device
difficult or not attachable
Speed control knob not working
High operating noise levels of product
Worn cutting edge of dermatome blade Replace dermatome blade
Performance of product is inadequate
Mechanical effect (e.g. dropped)
Remedy
Observe instructions for use (processing, care).
Preventative: lubricate product before every steril-
ization.
Have the manufacturer repair the product.
Follow operating instructions (operating mode).
Return the product to the manufacturer for repairs
Change tool.
Follow operating instructions (preparation, care).
Preventive measure: Oil product before every ster-
ilization.
Return the product to the manufacturer for repairs
Use the compatible lid for GA340/GA341.
Have the manufacturer repair the lid.
Have the manufacturer repair the product.
Use the compatible battery for GA340/GA341.
Have the manufacturer repair the battery.
Have the manufacturer repair the product
Use compatible sterile funnel or battery removal
device for GA340/GA341.
Change sterile funnel or battery device.
Have the manufacturer repair the product.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Return the product to the manufacturer for repairs
Have the manufacturer repair the product
Return the product to the manufacturer for repairs