7.11 Inspection, maintenance and checks

Allow the product to cool down to room temperature.
Note
Aesculap recommends lubricating the flap rod, flaps, sliding surfaces, and
other moving parts, (e.g. knob, cover lids) using STERILIT Power Systems oil
spray GB600 prior to every sterilization. Remove excess oil using a lint-free
wipe.
Inspect the product after each cleaning and disinfecting cycle for:
cleanliness, damage, function, abnormal operation noise, excessive
heat, or heavy vibration.
Set aside the product if it is damaged.

7.12 Packaging

Observe the instructions for use of the respective packaging and hold-
ers (e.g. Instructions for use TA009721 for Aesculap ECCOS holder sys-
tems).
Insert the products in the correct position into the ECCOS holders, see
Fig. G.
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
Ensure that the packaging will prevent a recontamination of the prod-
uct.

7.13 Steam sterilization

CAUTION
Damage or destruction of the batteries due to processing!
Do not sterilize the rechargeable battery.
Note
The product may only be sterilized with the blade, battery and lid as well as
opened blade cover all demounted.
9.

Troubleshooting list

Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Cause
Product not running
No battery
Battery not charged
Battery defective
Product in locked OFF position Safety catch is in OFF position
Product defective
Product becomes too hot Excessive use
Processing/care not per-
formed properly
Fall damage, product defec-
tive
Blunt tool
10
Note
The product can be sterilized with a mounted flap tab and flaps in the open
position.
Check to ensure that the sterilizing agent will come into contact with
all external and internal surfaces (e.g. by opening any valves and fau-
cets).
Use a validated sterilization method:
– Steam sterilization using fractional vacuum process
– Steam
DIN EN ISO 17665
– Sterilization using fractionated vacuum process at 134 °C/holding
time 5 min
When sterilizing multiple products in one steam sterilizer:
Ensure that the maximum permitted load specified by the manufac-
turer for the steam sterilizer is not exceeded.

7.14 Storage

Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
Store sterile single-use GB228R product in germ-proof packaging in a
dry, dark and temperature-controlled room.
8.

Maintenance

To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
Detection
No battery in the battery shaft
No signal tone when inserting the bat-
tery
No signal tone when inserting the bat-
tery
Product not running
Product becomes hot
Product heating
Product becomes hot
Heating of tool and product
sterilizer
DIN EN 285
and
Remedy
Insert battery.
Charge battery in charger.
Have the manufacturer repair the battery.
Turn the safety catch to the ON position.
Have the manufacturer repair the product.
Observe operating instructions (nominal operating
mode).
Follow operating instructions (preparation, care).
Preventive measure: Oil product before every ster-
ilization.
Return the product to the manufacturer for
repairs.
Change tool.
validated
pursuant
to