Herunterladen Inhalt Inhalt Diese Seite drucken

Mechanical Alkaline Cleaning And Thermal Disinfection; Inspection, Maintenance And Checks; Packaging - B. Braun Aesculap Acculan 4 GT666 Gebrauchsanweisung/Technische Beschreibung

Tplo säge
Vorschau ausblenden Andere Handbücher für Aesculap Acculan 4 GT666:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20
Aesculap
TPLO saw GT666
7.7.2

Mechanical alkaline cleaning and thermal disinfection

Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Insert the product in its proper position in the ECCOS fixation.
After automatic cleaning/disinfection, check visible surfaces for resi-
dues and repeat the cleaning/disinfection process as needed.
7.8

Inspection, maintenance and checks

Note
Aesculap recommends occasional spraying of moving parts (such as knobs,
couplings, lid caps) with the Aesculap STERILIT-Power-Systems oil spray.
Allow the product to cool down to room temperature.
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
Check the product for any damage, abnormal running noise, overheat-
ing or excessive vibration.
Inspect the tools/saw blades for broken, damaged or blunt cutting
edges.
Set aside the product if it is damaged.
12
Acculan 4
®
T
t
[°C/°F]
[min]
<25/77
3
55/131
10
>10/50
1
90/194
5
-
-
Water qual-
Chemistry/Note
ity
D–W
-
FD–W
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
0.5 % working solution
– pH ~ 11*
FD–W
-
FD–W
-
-
mind. 10 min at max. 120 °C
7.9

Packaging

Always follow the instructions for use of the respective packaging and
storage devices (e.g. instructions for use TA009721 for Aesculap-
ECCOS holder system).
Insert product 1 in the correct position in the ECCOS fixations, see
Fig. C.
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
Ensure that the packaging will prevent a recontamination of the prod-
uct.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis