3-3-1. Setting the number of divisions

The number of divisions (number of interpolations) is the number of times the wavelength for an input signal is divided. The
number of divisions (number of interpolations) that can be set varies depending on the interpolator model. The default settings
are 400 divisions for the MJ500, 1024 divisions for the MJ600, and 4000 divisions for the MJ700.
To change the number of divisions (number of interpolations), set mode switches 6, 7, and 8 as shown in the table below.
The maximum response frequency is the value for 1 Vp-p input and is intended for reference purposes only. The actual
maximum response frequency varies depending on the input conditions.
MODE switch 6
MODE switch 7
(setting switch 1 for
(setting switch 2 for
number of divisions)
number of divisions)
ON
OFF
ON
OFF
ON
OFF
ON
OFF
MJ500
MODE switch 8
(setting switch 3 for
number of divisions)
ON
ON
ON
ON
OFF
ON
OFF
ON
ON
OFF
ON
OFF
OFF
OFF
OFF
OFF
Number of divisions
Maximum response frequency
(number of
difference setting of 100 ns
interpolations)
400
360
300
240
200
120
100
80
at a minimum phase
20 kHz
22 kHz
25 kHz
32 kHz
42 kHz
70 kHz
70 kHz
70 kHz
(E) 9