Installation instructions
3.2.3
Certificates and approvals
Device certification FOUNDATION Fieldbus
The temperature field transmitter has successfully passed all the test procedures implemented
and has been certified and registered by the Fieldbus FOUNDATION. The device thus meets all
the requirements of the following specifications:
• Certified to fieldbus specification, revision status 6.1.2
• Device certification number: IT099000
• The device meets all the specifications of FOUNDATION Fieldbus-H1 (www.fieldbus.org)
• The device can also be operated with certified devices of other manufacturers
(interoperability)
3.3
Transport and storage
Carefully remove all the packaging material and protective covers that are part of the
transported package.
Dimensions and operating conditions: →  37
Pack the device so that it is reliably protected against impact when it is stored (and
transported). The original packaging offers the best protection.
Storage temperature
4
Installation instructions
If stable sensors are used, the device can be fitted directly to the sensor. For remote mounting
to a wall or stand pipe, two mounting brackets are available. The illuminated display can be
mounted in four different positions.
36
Without display –50 to +100 °C (–58 to +212 °F)
With display –50 to +80 °C (–58 to +176 °F)
iTEMP TMT162
Endress+Hauser