(Continued)
Voltage dips, voltage
variations and short
interruptions on power
supply input lines
IEC 61000-4-11*
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Note: Ut is the AC mains voltage before application of the test level.
*Applies to the optional electric lift and electric litter only.
Guidance and manufacturer's declaration - electromagnetic immunity
The Optional Electric Lift or Electric Litter and Optional Scale System is intended for use in the electromagnetic
environment specified below. The customer or user of the Optional Electric Lift or Electric Litter and Optional Scale
System should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
150 kHz to 80 MHz
Conducted RF IEC
61000- 4-6 *
6 Vrms in ISM radio
Radiated RF
IEC 61000-4-3
80 MHz to 2.7 GHz
www.stryker.com
EMC information
0% U
for 0.5 cycle at 0°,
T
45°, 90°, 135°, 180°, 225°,
270°, and 315°
0% U
for 1 cycle
T
70% U
(30% dip in UT) for
T
25 cycles
0% U
for 250 cycles
T
30 A/m
Compliance Level
3 Vrms
3 Vrms
6 V in ISM radio
bands
bands
3 V/m
3 V/m
1125-309-005 REV A
0% U
for 0.5 cycle at 0°,
T
45°, 90°, 135°, 180°, 225°,
270°, and 315°
0% U
for 1 cycle
T
70% U
(30% dip in UT) for
T
25 cycles
0% U
for 250 cycles
T
30 A/m
Electromagnetic Environment - Guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Optional Electric Lift or Electric
Litter and Optional Scale System, including
cables, than the recommended separation
distance calculated from the equation
appropriate for the frequency of the
Recommended Separation Distance
80 MHz to 800 MHz
800 MHz to 2.7 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the optional electric lift or
electric litter requires
continued operation during
power mains interruptions,
it is recommended that the
device be powered from an
uninterrupted power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
transmitter.
d=1,2√P
d=1,2√P
d=2,3√P
meters (m).
1-61