Manual cleaning/disinfection
Prior to manual disinfecting, allow water to drip off
for a sufficient length of time to prevent dilution of
the disinfecting solution.
After manual cleaning/disinfection, check visible
surfaces visually for residues.
Repeat the cleaning /disinfection process if neces-
sary.
Manual cleaning and wipe disinfecting
Phase
Step
I
Cleaning
II
Drying
III
Wipe disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water
quality at least)
RT:
Room temperature
Phase I
Clean the product under running faucet water,
using a suitable cleaning brush until all visible res-
idues have been removed from the surfaces.
Mobilize non-rigid components, such as set screws,
links, etc. during cleaning.
Phase II
Dry the product in the drying phase with suitable
equipment (e.g. cloth, compressed air), see Vali-
dated cleaning and disinfection procedure.
T
t
Conc.
[°C/°F]
[min]
[%]
RT
-
-
(cold)
RT
-
-
-
>1
-
RT
0.5
-
(cold)
RT
-
-
Water
Chemical
quality
D–W
-
-
-
-
Meliseptol HBV wipes 50 % Pro-
pan-1-ol
FD-W
-
-
-
Phase III
Wipe all surfaces of the product with a single-use
disinfectant wipe.
Phase IV
After the specified exposure time (at least 1 min),
rinse the disinfected surfaces under running FD
water.
Drain any remaining water fully.