Characteristic
Operating temperature
Operating humidity
Operating altitude
Characteristic
Power source voltage
Power source frequency
Power consumption
CAUTION. A ground connection through the power cord grounding conductor is
essential for safe operation.
Description
0 °C to +50 °C, with 5 °C/minute maximum
gradient, noncondensing, up to 3000 meter
altitude
5% to 95% relative humidity (% RH) up to
+30 °C
5% to 60% RH above +30 °C up to +50 °C,
noncondensing
Up to 3000 meters (9842 feet)
Description
Full Range: 100 to 240 VAC RMS ±10%,
Installation Category II (covers range of 90 to
264 VAC)
Sine wave crest factor must be between
1.30 and 1.41.
47 to 63 Hz over entire source voltage range.
360 to 440 Hz, 100 to 132 VAC RMS Source
Voltage.
All models: 80 W maximum