XRpad2 HWC-M Detectors User Manual
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
Do not modify or alter the X-ray detector, its components, or
accessories. Ignoring this warning may result in explosion, fire, or
electric shock, which may result in severe personal injury, death, or
substantial product damage.
The X-ray detector shall only be used with its approved Varex Imaging
accessories and replacement parts. Product certification and warranty
are rendered void if any alteration to the product is made, or any
instruction, warning, or caution is not followed. It is important that the
X-ray detector is not directly connected to the hospital (clinical)
network. Connection of the X-ray detector directly with the hospital
(clinical) network may disturb the IT environment and may influence
the effective data throughput. The imaging workstation and the
WLAN access point must comply with one of the following standards:
IEC 60601-1, IEC 61010-1, IEC 60950-1, or IEC 62368-1.
Connection of the X-ray detector to the hospital (clinical) network that
includes other equipment could result in previously unidentified risk
to patients, operators or third parties. The responsible organization
should identify, analyze, evaluate and control these risk. Subsequent
changes to the hospital (clinical) network could introduce new risks
and require additional analysis.
In wired mode the X-ray detector shall only be used in professional
healthcare facility environment.
In wireless mode the X-ray detector could be used in home
healthcare environment and professional healthcare facility
environment.
The system integrator shall consider the risks which may arise if the
X-ray detector is placed near to active HF SURGICAL EQUIPMENT
and the RF shielded room of an ME SYSTEM for magnetic resonance
imaging, where the intensity of EM DISTURBANCES is high.
Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the X-ray detector, including cables
specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
The X-ray detector is not designed to control X-ray dose. The system
integrator is responsible for controlling the X-ray radiation.