13.6

Maintaining the Battery Pack

To maintain the battery pack:
Before inserting the battery pack into the X-ray detector or battery charger, inspect the
battery pack for signs of damage, defects, or abnormalities. Do not use damaged, defective,
or abnormal conditioned battery pack.
Check to ensure the battery pack is clean, dry, and free of foreign contamination or debris.
If cleaning is necessary, see
instructions.
The battery pack has no repairable parts. Do not disassemble. No modification of this
product is allowed.
If the battery pack emits an odor or generates heat or in any way appears abnormal during
use, recharging, or storing, immediately remove it from the device or battery charger, and
stop using the battery pack.
Using a damaged or defective battery pack may reduce function time or cause the X-ray
detector system to fail.
If a battery pack leaks, do not touch the leaking fluid. If the fluid touches your skin or eyes,
wash the affected area with clean, running water and immediately seek medical attention.
If the battery pack has not been used or charged for an extended period of time
(approximately 30 days), check the condition of the battery pack, and recharge if necessary
before using.
13.7

Cleaning the Battery Pack

To clean the battery pack:
Avoid exposure of the battery pack to liquids and solvents when possible.
Do not allow liquids or solvents to contact the electrical contacts on the battery pack.
When necessary, clean the battery pack using a lightly moistened cloth with 70% isopropyl
alcohol or 3% hydrogen peroxide.
Never use thinner, benzene, acetone, or any other corrosive or flammable cleaning agents.
Ensure the battery pack is completely clean and dry before storing inserting the battery pack
into the X-ray detector, or inserting the battery pack into the battery charger.
48
Store and transport the battery in approved packaging only.
Retain the original packaging, for storing and transporting the
battery.