Scale electronic unit (NSW)
The following applies for the tare value:
(Tare Value)
Setting/reading the tare memory (ASCII format)
Remarks:
WE2108
If the "Manual tare" function is activated (Func menu, parameter F_tAr = 1), the manual
tare value is used on switching over to net.
Otherwise the value from earlier performed taring is used further.
TAV
Command:
Parameter:
Range:
Factory setting:
Password protection: no
Parameter backup:
Query:
All following net values are marked with the symbol "PT" (manual tare) independently of
the set taring function (Func menu / F_tAr).
In the "Manual tare" taring function (F_tAr = 1), the value entered with TAV is stored and
replaces the previous manual tare value.
After characteristic entries with the commands LDW, LWT (or SZA, SFA), the tare mem-
ory is set to 0.
The query TAV?; delivers the tare value which is momentarily used for calculating the net
value. This can be a manual tare value or the value of the last taring.
The tare value is scaled like the display and can also be entered optionally with decimal
point. The following notations are equivalent (example: display with 2 places after the
decimal point 000.00):
TAV2500;
TAV25.0;
TAV25.000;
TAV24.999;
TAV(yyyyy);
1
min. 1, max. [scale nominal value – 1]
0
with command TDD1
TAV?;
83