Scale electronic unit (NSW)
5
5.1
The procedure is analogous to the entries in the CAL menu. For calibration, the input signals
(between "2.7 mV / V) and the associated display value must be stored non−volatilely for
the different points of the characteristic.
Further information on partial load calibration, linearization and calculation of calibration val-
ues is provided in the operating instructions of the WE2108.
Effect of the user characteristic (exemplary values):
WE2108
Calibration commands
Overview
Internal characteristic, calibrated to mV / V
Interal digit
200000
140000
20000
I
0,2
Application measuring range
I : Calibration unloaded (LDW)
II: Calibration at nominal load (LWT)
User characteristic, determined by LDW / LWT
Display digit
6000
0
0,2
Application measuring range
Factory
characteristic
II
Load
1,4
2
in mV / V
Load
in mV / V
1,4
85