3 Care and maintenance
T3F Racer
59 48 968
T3F mini
59 48 976
28
•
Carefully insert the new turbine block into the cover.
•
Screw the cover together with the turbine block by hand
into the highspeed handpiece head and screw them
tight using the appropriate tool!
ATTENTION! Otherwise there is a risk that the cover
may come off and fall into the patient's pharynx together with
the turbine and the burr instrument.
•
Then lubricate (see pages 12–13).
•
Clamp in an FG instrument und stabilize the turbine block
in its O-ring seats by pulling and turning it.
•
Let the highspeed handpiece run briefly without cooling
spray.