Type of error
Compensation error:
An error has occurred in the compensation of influencing analysed readings.
A review of the compensation settings in the configuration of the analysis measuring input
in question is required.
Invalid display value
Possible errors are:
n
Error input: An erroneous signal is applied to an analogue input or there has been
selected an analogue input which is not equipped with an option board.
n
Error in the mathematics module formula: The result of a mathematical formula is
invalid (i.e. divided by zero)
Display overflow:
The display value is outside the limits -99999 to 99999.
85