7.
Ensure the power supply cord is placed on the oor under the bed. Improper placement of the
power supply cord could cause injury.
8.
Pull the coverlet over the pressure redistribution surface by stretching it over each corner securely.
Do not trap the power cord between the coverlet and mattress.
9.
Smooth any wrinkles on the coverlet.
10.
Secure the coverlet to the bed frame by using the hook-and-loop straps, located on the sides of the
coverlet.
11.
Ensure strap placement does not interfere with the operation of the bed functions. Failure to do so
could result in patient injury or equipment damage.
12.
Con rm there are no sharp objects in the immediate area which may damage the coverlet.
13.
Connect the power supply to an electrical outlet and con rm outlet has power. Verify that the
electrical outlet can be easily accessed when disconnecting the device from mains power. Ensure
power supply cord is properly stored on the oor beneath bed.
For Arjo surfaces with Skin IQ integration, the coverlet can be connected directly to the Skin IQ
port located on the pump via the fan assembly, removing the need to connect the coverlet to the
mains power. A separate accessory is required: Product Code – 636377 (Skin IQ Integration Coiled
Power Cable).
14.
Move hand along foot end of bed and feel for fan vibration. A low humming sound indicates the fan
is operating.
15.
To uninstall the Skin IQ 365 refer to the Care and Cleaning instructions.
Care and Cleaning
The Skin IQ 365 Coverlet should be cleaned regularly after each patient use. All sta members should
wear appropriate protective clothing when cleaning mattress. It is recommended that the top cover (patient
contact area) should be cleaned with a mild soap or organic solution and water prior to using standard
disinfection solutions in order to remove gross contaminates. All disinfection solutions must be properly
diluted according to manufacturer's instructions. This will be e ective on most stains including blood, urine
and perspiration. Please follow standard institutional cleaning and disinfection procedures.
The zippered area must not be opened during normal use or during cleaning or disinfection
procedures.
Disinfection
The Skin IQ 365 Coverlet, Fan and Power Supply can be disinfected using a wipe down approach with
a 70% alcohol solution, bleach 1000 ppm, hydrogen peroxide 3%, or phenolic disinfectant 1%. Although
bleach may be used for wipe-down disinfection, please be aware that if used extensively, discoloration
and fading of the fabrics and labels used in the Skin IQ 365 may occur. Bleach should not be used when
laundering the Skin IQ 365.
Disinfection solution e cacy data was based on microbiological studies using Staphylococcus aureus as a
common clinically relevant aerobic skin infection organism.
Iodophor type disinfectants (e.g. Betadine, etc.) are not recommended and will stain fabric.
The zippered area must not be opened during normal use or during cleaning or disinfection
procedures.
Laundering
Prior to laundering unplug the power supply from the fan assembly and remove the fan
assembly from the red port on the Skin IQ 365 coverlet.
Do not bleach when laundering the Skin IQ 365.
Laundering of the Skin IQ 365 is permitted. Please follow standard institutional laundering procedures. It is
recommended that wash temperatures do not exceed 140 F (60 C) and drying temperatures do not exceed
140 F (60 C). Be sure to unzip the Skin IQ 365 when washing and drying to allow the inside of the coverlet
to be cleaned, disinfected and dried. When laundering process is complete, be sure to re-zip the Skin IQ
365 before it is returned to use. A tag located on the foot end of the Skin IQ 365 is used to track the number
of times the coverlet is laundered.
Temperatures in excess of what is recommended will cause premature deterioration of the
Skin IQ 365 coverlet.
The zippered area must be opened during laundering procedures, but closed during normal
use.
End of Life Disposal
The coverlet itself is a multi-patient reusable product, but some of the items that come with it can be reused
after the products expected service life, if they are handled properly when removed. The Skin IQ 365 power
supply can be reused, but only on other Skin IQ 365 fans. Routinely inspect the power supply and cord and
remove from service if damaged. Follow the steps below to remove the components of the Skin IQ 365.
1.
Disconnect power supply from wall outlet.
2.
Disconnect power supply from fan.
Consider all facility policies and procedures with regard to cleaning, inspection and reuse of
electronic equipment. If not reused, dispose of power supply per approved local institutional
protocols.
3.
Remove fan from coverlet by pulling fan apart from the coverlet.
4.
Dispose of coverlet and fan according to approved local institutional procedures.
Fabric material used on the coverlet or any other textiles, polymers or plastic materials etc. should be sorted
as combustible waste.
Units have electrical and electronic components that should be disassembled and recycled per Waste of
Electrical and Electronic Equipment (WEEE) or in accordance with local or national regulation.
Fan assembly contains electronic components that may require alternate disposal than the
soft goods of the coverlet.
Improper disposal of any component may result in regulatory non-compliance.
Speci cations
Speci cations subject to change without notice.
Maximum Weight Capacity ............................................................................................... 500 lb (227 kg)
Useful life of this product is one year or 35 laundry cycles, whichever comes rst.
Consult the speci cations for the pressure redistribution surface being used. Additional weight
limitations may apply.
Electrical:
Voltage Input............................................................................................................................... 230 VAC
Voltage Output ................................................................................................................................... 12 V
Frequency........................................................................................................................................ 60 Hz
Ampere Rating Input...................................................................................................0.3 A (0.3 A - 0.1 A)
Ampere Rating Output ...................................................................................................................... 1.0 A
Maximum Electrical Leakage........................... 100 uA at 115 VAC 60 Hz and 200 uA at 230 VAC 50 Hz
Power Cord Length............................................................................................................. 20.0 ft (6.1 m)
Environmental Conditions:
Operating:
Temperature Range ........................................................................................ 32 F (0 C) to 104 F (40 C)
Humidity Range ........................................................................................................................ 5% - 90%
Transport / Storage:
Temperature Range ................................................................................ -20.2 F (-29 C) to 167 F (75 C)
Humidity Range ........................................................................................................................ 5% - 95%
The Skin IQ 365 Coverlet is classi ed as a Type B applied part under IEC 60601-1.
Electromagnetic Compatibility
Electromagnetic Interference - Although this equipment conforms with the intent of the directive 2014/30/ EU
in relation to Electromagnetic Compatibility (EMC), all electrical equipment may produce interference. If
interference is suspected, move equipment away from sensitive devices or contact manufacturer.
Portable and mobile Radio Frequency (RF) communications equipment can e ect electrical equipment.
Radios, cell phones and similar devices may a ect this equipment and should be kept at least 6.5 ft (2 m)
away from the equipment.
The Skin IQ 365 uses RF energy only for its internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electrical equipment. The Skin IQ 365 is intended for use
in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the
user of the Skin IQ 365 can help prevent electromagnetic interference by maintaining a minimum distance
between the portable and mobile RF communications equipment (transmitters) and the Skin IQ 365.
Symbols Used
Conforms to AAMI ES60601-1-6.
Certi ed to CSA Std. C22.2 No. IEC-60601-1 edition 3.1
IEC-60601-1-2:2014
5021463
Important Operational Information
Foot End
CE marking indicating conformity with European
Community harmonised legislation
Figures indicate Noti ed Body supervision.
Consult Instructions for Use
Indicates the product is a Medical Device according
to EU Medical Device Regulation 2017/745
Class II Device
Manufacturer
Temperature Limitations
REF
IPX1
Protected against ingress of liquids
Hospital name
60 C
Washable up to 60 C
60 C
Do not iron
Customer Contact Information
For questions regarding this product, supplies, maintenance or additional information about Arjo products
and services, please contact Arjo or an Arjo authorized representative or visit www.arjo.com. In the US call
1-800-343-0974.
Design Policy and Copyright
and
are trademarks belonging to the Arjo group of companies.
®
™
© Arjo 2022 .
As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice. The content of
this publication may not be copied either whole or in part without the consent of Arjo.
At Arjo, we are committed to improving the everyday lives of people a ected by reduced mobility and age-related
health challenges. With products and solutions that ensure ergonomic patient handling, personal hygiene, disinfection,
diagnostics, and the e ective prevention of pressure ulcers and venous thromboembolism, we help professionals across
care environments to continually raise the standard of safe and digni ed care. Everything we do, we do with people in
mind.
ArjoHuntleigh AB
Hans Michelsensgatan 10
211 20 Malmö, Sweden
www.arjo.com
Date of rst use
Do Not Open With Scissors
Tripping Hazard
This product or its parts are designated
for separate collection at an appropriate
collection point. At the end of useful service
life, dispose of all waste according to local
requirements, or contact your local Arjo
representative for advice.
Keep Dry
Warning of possible hazard to system,
patient or sta
No Hooks
Do Not Shower
Type B Applied Part
Catalog Number
Humidity Limitations
Tumble dry at 60 C