3.2. Tubing Set-up.

WARNING: This product is for single-use only. It has not been designed to be re-used/re-sterilized.
Reprocessing may lead to changes in material characteristics such as deformation and material
degradation which may impact the strength of the device and compromise device performance.
Reprocessing of single use devices can also cause cross-contamination leading to patient infection.
These risks may potentially affect patient safety.
This product contains DEHP, a phthalate chemical known to cause birth defects and other reproductive
harm, including in children and in pregnant or nursing women.
Sterility is preserved only if the tube changing procedures have been performed precisely according to the
set-up instructions below and in Section 3.2.E.
A. Installing the One Day Set Irrigation
Tube Ref. 284503 or 284504.
Contact your DePuy Mitek Representative for
ordering information since certain tubes are not
offered in all countries.
(A)
Clamps
From
solution
bags
Clamps
(A)
Close all white clamps and
connect to fluid bags.
This tubing connects the saline solution bags to
the irrigation tubing. Solution flows from the spikes
(A) around the roller pump, through the pressure
chamber, where the pressure is read through a
filter (B) by a built-in pressure transducer.
GREEN
Auto-locking device
(B)
To the Pressure
Filter
transducer
Slide the green part of the
tube into the auto-locking
device on the pump.
20
PRESSURE CHAMBER
Clamps
Place and center the tube
around the roller pump.
(C)
To Patient Set
Irrigation Tube