Filling the reservoir with sterile distilled water (Continued)
2.
Fill the reservoir with five liters of sterile distilled water. Do not fill past the top fill line to avoid water overflow
(Figure 1-17 on page
1-29).
Selecting and setting the primary probe
A patient probe displays when present, stable, and confirmed. The choice of Probe A or Probe B is highlighted when you
insert the cable into port A and port B. If you only insert one cable, the active port displays.
1.
Tap the Settings button.
a.
In the standby mode, tap the Back button to display the edit settings screen.
b.
In an active therapy mode, tap the Next button.
2.
Tap Select Probe to display the Select Primary Probe (Probe A or Probe B) screen. If both probes are present, the
default is probe A.
3.
Tap A or B, if applicable.
4.
Tap Confirm.
Notes
•
The message "Probe stabilization in progress... Please wait" is displayed.
•
When you initially select a probe (A or B), detected is checked. When stabilized, the Ready check is displayed.
•
If the probe is not stabilized within three minutes, the message "Probe stabilization error" will appear. Tap the
Help button for more detail.
•
You can select help at any time to display help with the current screen or icon descriptions.
Filling a thermal transfer device
Read the operations manual for the individual thermal transfer devices for warnings, cautions, and safe operating
instructions before use.
Note: These instructions are to pre-fill the thermal transfer devices only, not therapy. See
33.
To fill a thermal transfer device:
1.
Connect a thermal transfer device following:
www.stryker.com
Operation
A
Figure 1-17: Reservoir fill lines
Connecting and disconnecting thermal transfer devices on page
8001-109-005 REV B
Switching modes on page 1-
1-25.
1-29
English
EN