English
Product alarms (Continued)
EN
Icon, yellow
1-18
Introduction
Name
Alarm priority
and delay
Probe or adapter
Medium, 30
malfunction (A or
second delay
B)
Adapter cable
Medium, 30
disconnected (A
second delay
or B)
Patient
Medium
temperature
deviation
Patient
Low
temperature
deviation
Normothermia
Low
deviation
Therapy pause
Medium
Battery low
Low
Patient
Low
temperature
output (A or B)
Message
No temperature
signal detected
Adapter cable
not detected
Patient
temperature is
more than ± 0.5°
C from target
temperature
(Only will appear
after the initial
target is reached.)
Patient
temperature is
more than ± 0.5°
C from target
temperature
(Only will appear
after the initial
target is reached.)
Patient
temperature is
outside of 36.0°
C (96.8° F) to
38.0° C (100.4° F)
Therapy is
currently paused
Battery is low
Patient
temperature
output is
inaccurate on the
external device,
or outside
supported range
8001-109-005 REV B
Therapy
Check
interrupted
Yes
Check probe or
adapter cable
condition, location,
and connections
Yes
Reinsert the adapter
cable. If damaged,
replace the adapter
cable
No
Check patient
condition,
placement of
thermal transfer
devices, and all
connections
No
Check patient
condition,
placement of
thermal transfer
devices, and all
connections
No
Check patient
condition,
placement of
thermal transfer
devices, and all
connections
Yes
To resume, press
and hold the play /
pause for 2 seconds
No
Maintenance is
recommended. If
battery is not
replaced, the
product may not
function on the next
startup.
No
Check output
adapter cable
connection. Tap
Confirm to
reactivate the output
port.