11. 1 IRRIGATION LINE MAINTENANCE

Sterilizable irrigation line
Refer to the user's manual delivered with the
equipment.
The irrigation line delivered with the I
unit are sterilizable. The perforaters are single-use
and delivered sterile.
Sterile irrigation line
Satelec can provide disposable irrigation lines which
must always be discarded, in safety container for
soiled medical items, after use. Re-use of a single-use
irrigation line may result in contamination of your
patients and expose you to de facto liability. Do not
attempt to resterilize the single-use irrigation lines.
Do not attempt to modify the irrigation lines.

11. 2 CONTRA-ANGLE MAINTENANCE

Refer to the manufacturer's instructions for your
contra-angle.
11. 3 ROTARY INSTRUMENT MAINTENANCE
Please refer to the manufacturer's instructions for
your instruments (burs, drills, etc.).

11. 4 DEVICE MAINTENANCE

Important:
Never use an abrasive cleaning agent on the
device.
Do not use sprays or liquids to clean and disinfect
the I
C
2 control unit.
MPLANT
ENTER
The device, the brackets and the multifunction
footswitch are not sterilizable.
The I
C
MPLANT
ENTER
footswitch and the bracket must always be cleaned
and disinfected after each procedure, using alcohol,
disinfectant agents or disinfectant wipes for dental
practices, such as S
EPTOL
Regular monitoring of the I
necessary in order to detect any problem.
20
MPLANT
2 control unit, the control
™ wipes.
C
MPLANT
ENTER
It is important to keep the control unit ventilation
vents clean in order to avoid abnormal heating.
11. 5 MICROMOTOR CORD AND SCALER CORD
MAINTENANCE
C
2
ENTER
Important:
Never use an abrasive cleaning agent on the cords.
The cords must not be immersed or disinfected
using agents containing acetone, chlorine or
bleach.
The cords must not be cleaned in an ultrasonic
bath.
a) Cleaning and disinfection
The cords must always be cleaned, disinfected and
sterilized after each procedure, using alcohol,
disinfectant agents or disinfectant wipes for dental
practices, such as S
b) Sterilization
The cords can be sterilized in an autoclave under the
following conditions:
- Autoclave: Class B
- Sterilization temperature: 134°C at 2 Bar.
- Time at sterilization temperature: 18 minutes.
CAUTION: Not all autoclaves can reach 134°C. Not
all autoclaves draw a pre-vacuum. Please refer to
your
autoclave
sterilization instructions.
11. 6 I-SURGE LED MICRO-MOTOR MAINTENANCE
Refer to the user's manual delivered with the
equipment.

11. 7 ULTRASONIC HANDPIECE MAINTENANCE

After each use, the handpiece irrigation circuit must
be rinsed with distilled or demineralized water for 20
to 30 seconds.
Disconnect the N
2 unit is
before cleaning, disinfection and sterilization.
™ wipes.
EPTOL
manufacturer
L
handpiece from its cord
EWTRON
ED
for
specific