VI - MAINTENANCE AND
STERILIZATION
The following maintenance and/or sterilization
instructions should be carried out before using
the accessories supplied with the device for the
first time.
Avoid using cleaning agents and disinfectants
containing inflammable substances. If this is
unavoidable, ensure that the product has
evaporated and there is no combustible product
on the device and accessories before operation.
6. 1 SUPRASSON
NEWTRON
MAINTENANCE
After each use, the water circuits of the whole
unit comprising the S
UPRASSON
should be rinsed with clean water for 20 to 30
seconds.
For sterilization of the handpiece, please refer to
the specific instructions for the product.

6. 2 TIP MAINTENANCE

a) Life
The shape and the mass of the tips are the
determining factors for obtaining the best results
from the ultrasonic generator. Vigilance on the
part of the user concerning these two
characteristics ensures the device's consistent
performance.
12
In consequence, we strongly advise against
modifying the structure of the tips by filing or
bending them.
Similarly, the ageing of a tip through normal wear
leads to changes in its characteristics.
Always replace a tip that is worn or which has
suffered an accidental impact (fall etc.).
Use the TIPCARD to check the wear of scaling
tips. Renew the most frequently used tips a
minimum of one to three times a year.
b) Sterilizing tips
All tips should be cleaned before sterilization
with cotton wool soaked in alcohol, wipes or by
immersion in an ultrasonic tank.
Sterilizing tips can be carried out by all
conventional heat processes, the sterilization
conditions and cycles depending on the
HANDPIECE
equipment.
®
Sterilizing equipment: autoclave
Sterilization under wet heat - instrument
wrapped: 134°C at 2 bar (200 KPa) for 18
N
and the tip
minutes.
®
EWTRON
WARNING: during sterilization, metal parts of
different types must not touch each other. Any
contact will lead to the formation of
electrolytic
deterioration. In order to avoid this happening,
place the objects to be sterilized in a cloth or
sterilizing bag.
couples
causing
localized