- Prime the irrigation circuit by pressing the
flush/prime button on the screen (Fig. 3-1) or the
multifunction footswitch (Fig. 11-2).
- When the irrigation solution reaches the end of the
tip, release the footswitch.
- Use the I
C
MPLANT
ENTER
dental practice.
X - SHUTTING DOWN THE
DEVICE
Upon completion of the dental procedure:
- Remove the bottles or bags of physiological saline or
sterile water from the brackets.
- Remove the irrigation line perforating pins from the
bottles or bags.
- Immerse the irrigation line perforating pins in a
vessel containing distilled water.
- Rinse each irrigation line alternately, along with the
contra-angle and the handpiece, by running the
flush/prime function until the vessel and the
irrigation lines are completely empty.
- Remove the irrigation line clips.
- Disconnect the irrigation lines from the contra-
angle and from the handpieces.
- Remove the rotary instruments fastened to the
contra-angle and the tip screwed onto the
handpiece.
- Disconnect the contra-angle from the I-S
micro-motor.
- Disconnect the micro-motor cord from the
I-S
L
.
URGE
ED
- Disconnect the P
IEZOTOME
N
L
handpiece and the scaler cord.
EWTRON
ED
- Switch the device OFF (O).
2 in accordance with good
URGE
2 L
handpiece or the
ED
XI - ROUTINE MAINTENANCE /
STERILIZATION
Important:
The device must be switched OFF during cleaning
and disinfection procedures.
The only irrigation lines that can be sterilized are
those identified as sterilizable.
During sterilization, items made of different metals
must not come into contact. Any contact would
establish galvanic couples and cause localized
damage.
To avoid this happening, place the items in
individual sterilizable bags or in a sterilization box.
To maintain the sterility or asepsis of the
accessories (contra-angle, micromotor, micromotor
cord, cords, handpieces, etc.), store them in
sealed bags or containers suitable for medical
practice.
The maintenance and/or sterilization instructions
below must be applied before each use of the
device.
Avoid using cleaning and disinfection agents
containing flammable substances.
If this is not possible, make sure that all the agent
L
has evaporated and that there is no combustible
ED
matter on the device or its accessories before it is
switched on.
Sterilized items must be allowed to dry and cool to
room temperature before they are reused.
Before sterilization, check that your autoclave is
clean and that the water used is of adequate
quality.
After each sterilization cycle, remove the items
from the autoclave immediately in order to
minimize the risk of metal corrosion.
19