- Plug the handpiece into the handpiece cord.
- Remove the irrigation tank covers.
- Fill the tanks with irrigation solution without
exceeding the level indicated by the borders
(Fig. 4-1).
- Replace the irrigation tank covers.
- Install the filled tanks on their supports.
- Switch on the device (I)
- Set the device to maximum power (1).
- Prime the irrigation circuit by pressing the flow
setting button (Fig. 1-6).
- When the irrigation solution reaches the end of
the handpiece, release the button.
- Install the required tip on the handpiece.
- Moderately tighten the tip using the wrench to
ensure optimal ultrasonic operation.
- Set the device to the required power (according
to color and number) using the control dial (Fig.
1-4).
- If necessary, refer to the TipBook to find out
SATELEC tip usage power settings.
- Select the type of irrigation required (liquid
irrigation/no irrigation/air).
Liquid irrigation:
• Adjust the desired irrigation flow rate.
• Select the irrigation solution tank.
• Press the "Irrigation ON/OFF" button on the
multifunction footswitch (Figure 3, (3)).
• Press the footswitch to use your device.
No irrigation:
• Neither irrigation solution tank should be on.
• The "air" indicator should not be on.
• If any functions have been activated, press the
button of the particular function (irrigation or
air)
• Press the footswitch to use your device.
12
"Air" irrigation:
• Press the "air" button on the multifunction
footswitch (Fig. 3, (4)).
• The "air" indicator must be switched on.
• Press the footswitch to use your device.
VII - SHUTTING DOWN THE
DEVICE
At the end of the working day, you must:
- Empty the irrigation solution out of the
irrigation circuit and the irrigation solution
tanks, by pressing the irrigation flow setting
button (Fig. 1, item 6).
- Rinse the irrigation circuit and the irrigation
tanks (see chapter 4, (4.4)) with distilled or
demineralized water.
- Flush on air to dry the irrigation circuit.
- Disconnect the handpiece cord from the
handpiece.
- Disconnect the handpiece cord from the device.
- Switch the device off (O).
VIII - ROUTINE
MAINTENANCE/
STERILIZATION
Important:
To maintain the sterility or asepsis of the
accessories (handpiece, tips, etc.), store them
in sealed bags or leakproof containers suitable
for medical practice.