799 GPT Titrino
3.2.2.40. Mode.Parameter.CtrlPara.MinRate
Parameter for STAT: Control parameters.
.MinRate: Minimum titration rate in ul/min.
3.2.2.41. Mode.Parameter.TitrPara.TStart
Mode.Parameter.TitrPara.StartMeas
Mode.Parameter.TitrPara.RateStart
Parameters for STAT: Start of the measured point acquisition after the start
condition has been met.
.TStart: Start time in s.
.StartMeas: Start measured value.
.RateStart: Start rate in mL/min. The current rate must be less than the start
rate .This criterion does not become effective until 10 s after the
start.
3.2.2.42. Mode.Parameter.StopCond.TStop.Type
Mode.Parameter.StopCond.TStop.Time
Mode.Parameter.StopCond.TStop.Factor
Mode.Parameter.StopCond.TStop.Delta
Mode.Parameter.StopCond.TStop.Delay
Parameters for STAT: Stop time.
If an absolute stop time (abs.) has been selected, the time is in s (Time).
A relative stop time (rel.) is calculated as a function of the sample size:
Stop time in s = sample size * factor
The factor selected (Factor) applies.
delta: Time in s after control point is reached for the first time.
delay: Time in s after last dosing step.
OFF means that the criterion is not monitored.
3.2.2.43. Mode.Parameter.StopCond.RateStop
Parameters for STAT: Stop rate in ml/min. The current rate must be less than
the start rate. This criterion does not become effective until 10 s after the
start.
3.2.2.44. Mode.Parameter.Evaluation.TimeWin.1.LowLim 0...999 999, OFF
Mode.Parameter.Evaluation.TimeWin.1.UpLim
Parameters for STAT: Evaluation of rates in time windows. Entry of the times
in s. The evaluated rates are available as variables C8X (X = 1...9).
3.2.2.45. Mode.Parameter.Evaluation.FixVol.1.Value
Parameters for STAT: Interpolation of the volume at preset times. Entry of the
times in s. The interpolated volumes are available as variables C5X.
3.2 Remote control commands
0.01...25.0...9999
pH: 0.00...±20.00, OFF
U, Ipol: 0...±2000, OFF
Upol: 0.0...±200.0, OFF
0.01...150, OFF
abs.,rel.,delta,delay,OFF
0.01...150, OFF
0...999 999, OFF
0...999 999, OFF
0...999 999
0...999 999
0...±999 99
0...999 999
0...999 999
151