8.
Troubleshooting list
Malfunction
Dermatome blade not
moving.
Insufficient cutting per-
formance of dermatome
blade
Cutting depth
uneven/deviating from
setting
9.
Technical Service
Danger to life of patients and
users if the product malfunc-
tions and/or protective mea-
WARNING
sures fail or are not used!
For service and repairs, please contact your
national B. Braun/Aesculapp agency.
Modifications carried out on medical technical equip-
ment may result in loss of guarantee/warranty rights
and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the
address indicated above.
Detection
Loud running noise from
motor
Worn cutting edge of
dermatome blade
Insufficient motor
power/speed
Skin grafts uneven / too
thick / too thin
Do not perform any servic-
ing or maintenance work
under any circumstances
while the product is being
used on a patient.
Do not modify the product.
Cause
Defective gearings
Dermatome blade
blunted
Battery too weak
Battery worn/defective
Adjusted dermatome
Mechanical influence
(e.g. fall on floor)
Remedy
Have repaired by manufac-
turer
Replace dermatome blade
Charge battery
Replace battery
Have repaired by manufac-
turer
Have repaired by manufac-
turer
en
21