Neoss NeoScan 1000 Bedienungsanleitung Seite 7

Vorschau ausblenden Andere Handbücher für NeoScan 1000:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 17
Computer:
• DO NOT place any equipment which does not comply with IEC 60601-1 in the immediate vicinity of the patient.
Leave at least 1.5 meters distance between the patient and the equipment.
• The scanner is only intended to be connected to a computer that is at least IEC 60950 / IEC 62368, or
equivalent standards certified. Connecting the scanner to other equipment may be hazardous.
• See the installation guide for your computer for information about the data processing system, computer, and
screen. Leave a sufficient amount of clear space around the computer to ensure that it is properly ventilated.
• Position the screen to avoid light reflections from internal or external lighting for maximum image quality and
visual comfort.
Disposal:
This equipment contains certain materials and chemical compounds incidental to the manufacture of
electrical and electronic equipment, and improper "end-of-life" disposal of such equipment can result in
environmental contamination. Therefore, this equipment should not be disposed of as ordinary household
waste but should instead be delivered to a designated electrical and electronic waste disposal or
recycling center. For further information on disposing of electrical and electronic waste, contact the
cognizant authority within the local jurisdiction. Dispose of the scanner tips according to standard
operating procedures or local regulations for the disposal of contaminated medical waste. For additional
scanner tips, contact your dealer.
Cleaning, disinfecting, sterilizing
Perform the following maintenance activities on your scanner and accessories regularly.
To ensure maximum hygienic safety for the patient, carefully follow the instructions to prepare the scanner for use.
To ensure maximum hygienic safety for the patient and to minimize the risk of cross-contamination, carefully
perform the following maintenance activities on your scanner and accessories.
After each patient:
• Clean and disinfect the scanner. See "Clean and Disinfect the Scanner".
• Reprocess the tip by applicable method:
– Clean the tip, and then perform High Level Disinfection (See "Clean the Tip" and "Disinfect the Tip").
– Clean the tip, and then perform Autoclave Sterilization (See "Clean the Tip" and "Sterilize the Tip").
There are 3 models of tip as following, applicable to different reprocessing method:
Model
Size
TP101N
Large
TP102N
Large
TP103N
Small
Figure
Cleaning
(01)06973993441072
(01)06973993441058
(01)06973993441065
Manual
Automatic
Cleaning
High Level
134°C
Disinfection
Sterilization
x
x
x
NeoScan 1000 // Safety information
7