Guidance and Manufacturer's declaration – Electromagnetic immunity for equipment and systems
The NeoScan 1000 is intended for use in the electromagnetic environment specified below. The customer or the
user of the NeoScan 1000 should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Radiated RF
IEC 61000-4-3
For the immunity to proximity fields from RF wireless communications equipment, the NeoScan 1000 is
compliant with the test levels specified below, according to IEC 60601-1-2 standard. The customer or user of
the NeoScan 1000 should assure that it is used in such an environment.
Test Frequency
(MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
14
NeoScan 1000 // Regulatory information
IEC 60601 Test Level
Compliance Level
±8 kV contact
±15 kV air
30 A/m
3 V/m
80MHz – 2.7GHz
Band
(MHz)
380 – 390
Pulse modulation 18Hz, 27V/m
430 – 470
FM, ±5 kHz deviation, 1 kHz sine, 28V/m
704 – 787
Pulse modulation 217Hz, 9V/m
800 – 960
Pulse modulation 18Hz, 28V/m
1700 – 1990
Pulse modulation 217Hz, 28V/m
2400 – 2570
Pulse modulation 217Hz, 28V/m
5100 – 5800
Pulse modulation 217Hz, 9V/m
Electromagnetic Environment – Guidance
±8 kV contact
Floors should be wood, concrete, or ceramic
±15 kV air
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
30 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
3 V/m
Environment of a professional healthcare
80MHz – 2.7GHz
facility.
WARNING: Portable RF communications
equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to
any part of the NeoScan 1000 including cables
specified by the manufacturer. Otherwise,
degradation of the performance of this
equipment could result.
Immunity Test Levels