11 Percent weighing
The application Percent weighing allows to check the weight of a sample in percent,
referred to a reference weight.
• If you use a weighing container, it should be tared before setting the
reference quantity (see chap. 9.3)
• The readability of the balance automatically adjusts to the reference
weight:
Readability in %
Model
TPES 620-3-B
TPES 2200-2-B
TPES 4200-2-B
TPES 6200-2-B
TPES 15000-1-B
TPES 31000-1-B
TPEJ 620-3M-B
TPEJ 2200-2M-B
TPEJ 4200-2M-B
The reference weight can be recorded in two ways:
• Actual value setting method: Weighing the reference weight
• Numeric entering of the reference weight
34
Weight range of the reference weight
1
Minimum load <= Reference weight < Minimum load x 10
0.1
Minimum load x 10 <= Reference weight < Minimum load x 100
0.01 Minimum load x 100 <= Reference weight
Minimum load for percent weighing
Select
(selection see chap. 9.4)
The display shows <%>.
Press and hold the [F]- key for about 2
seconds.
0.1 g
1 g
10 g
0.1 g
1 g
1 g
weighing
mode
<1.
TPES-B_TPEJ-B-BA-def-2319
SEt
3>