a Indications for Use
The Ranger blood/fluid warmer is intended to warm blood, blood products,
and liquids.
Patient Population and Settings
Adult and pediatric patients being treated in operating rooms, emergency
trauma settings, or other areas where blood/fluids are infused.

Explanation of Signal Word Consequences

WARNING: Indicates a hazardous situation which, if not avoided, could
result in death or serious injury.
CAUTION: Indicates a hazardous situation which, if not avoided, could result
in minor or moderate injury.
WARNING: To reduce the risks associated with air embolism:
• Never infuse fluids if air bubbles are present in the fluid line
• Ensure all luer connections are tightened
• Do not mount the Ranger blood fluid warming system higher than patient
level, as air embolism may result
WARNING: To reduce the risks of biohazard exposure, cross‑contamination
or infection:
• Do not reuse. This product is designed for single patient use only.
• Do not use if caps are not in place
• Ensure that all luer connections are securely tightened prior to priming set
• Do not exceed 300 mmHg pressure; ensure inline hand pressure pumps or
other pumps do not exceed 300 mm/Hg pressure.
WARNING: To reduce the risk of leaking and loss of therapy:
• Do not reuse. This product is designed for single patient use only.
• Ensure that all luer connections are securely tightened prior to priming set.
WARNING: To reduce the risks associated with potential blood loss:
• Do not use in combination with an extracorporeal circuit.
WARNING: To reduce the risk associated with patient exposure to leakage
current that exceeds Type BF equipment limits:
• Do not use in combination with other medical equipment without verifying
that the total leakage current from the combined equipment does not exceed
the safety limits for Type BF equipment.
CAUTION: To reduce the risk of leaking and loss of therapy:
• Do not exceed 300 mmHg pressure; ensure inline hand pressure pumps or
other pumps do not exceed 300 mm/Hg pressure.
• Do not remove warming set from warming unit while in use.
CAUTION: To reduce the risks associated with warming platelets,
cryoprecipitate, or granulocyte suspensions:
D
C
B
E
F
G
M
• Follow the AABB Guidelines for the use of blood warming devices which
caution against warming when administering platelets, cryoprecipitate, or
granulocyte suspensions.
CAUTION: To reduce the risks associated with potential material toxicity:
• Do not use in combination with an extracorporeal circuit as the Ranger
blood/fluid warming system is not designed for use with a circulating blood
system; therefore, biocompatibility testing has not been completed to
assess genotoxicity.
Instructions for Use
Do not use if package has been previously opened or damaged. Refer to the
3M™ Ranger™ blood/fluid warming unit model 245 Operator's Manual for set
up and use of the warming system.