Aesculap Spine
activ C insertion instrument
Manual pre-cleaning with a brush
Phase
Step
I
Disinfectant
cleaning
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
Phase I
Completely immerse the instrument in the cleaner/
disinfectant solution. Ensure that all accessible
surfaces are moistened.
Clean the product with a suitable cleaning brush
until all discernable residues have been removed.
Use a suitable cleaning brush to clean all surfaces
which are not visible, e.g. in instruments with
concealed crevices, lumens or complex geometries,
for at least 1 min or until no further residues can
be removed.
Maneuver
non-rigid
adjustable screws, hinges etc during cleaning.
After cleaning, thoroughly rinse through these
components (at least five times) with the cleaning
solution, using a disposable syringe (20 ml).
Do not use metal cleaning brushes or other
abrasives that would damage the product surfaces
and could cause corrosion.
Phase II
Rinse/flush
the
instrument
accessible surfaces) under running water.
10
T
t
Conc.
[°C/°F]
[min]
[%]
RT
15
(cold)
RT
1
(cold)
components,
such
as
thoroughly
(all
Water
quality
2
D–W
BBraun Stabimed; aldehyde-free,
phenol-free and QAV-free
-
D–W
-
Chemical