Aesculap
®
S4-Spondylolisthesis reduction instrument (SRI) FW225R
Validated cleaning and disinfection procedure
Validated procedure
Manual pre-cleaning
with ultrasound, brush
and water pistol and
subsequent mechanical
alkaline cleaning and
thermal disinfection
Mechanical cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must be of tested
and approved effectiveness (e.g. FDA approval or CE
mark according to DIN EN ISO 15883).
Manual pre-cleaning with ultrasound, brush and water pistol
Phase
Step
I
Ultrasound cleaning
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed
►
Observe information on suitable cleaning brushes
and water pistol, see Validated cleaning and disin-
fection procedure.
8
Specific requirements
■
Suitable cleaning brush
■
Water pistol: 3 bar to 4 bar
■
For products with movable joints,
move the joints while cleaning.
■
Connect components with lumens
and canals directly to the special
rinsing port of the injector cart.
■
Place the product in a tray that is
suitable for cleaning (avoiding
rinsing blind spots).
D
t
[°C/°F]
[min]
RT (cold)
>5
RT (cold)
1
Reference
Chapter Mechanical cleaning/disinfection
with manual pre-cleaning and sub-chap-
ter:
■
Chapter Manual pre-cleaning with
ultrasound, brush and water pistol
■
Chapter Mechanical alkaline cleaning
and thermal disinfection
Note
The cleaning and disinfection device used for process-
ing must be serviced and checked at regular intervals.
Conc.
Water
Chemistry
[%]
quality
2
D–W
1 % BBraun Cleaner N
-
D–W
-