Herunterladen Inhalt Inhalt Diese Seite drucken

To Verify/Record The Act Plus™ Actuator Heat Block Temperature; To View The Due Status For Quality Control Testing; Qc Manager Functions - Medtronic ACT Plus Benutzerhandbuch

Automatisches gerinnungszeit-messgerät
Vorschau ausblenden Andere Handbücher für ACT Plus:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

ACT_CH.fm 6/23/04 1:23 pm
UC200304151c ML
5.5" x 7.75" (139.7 mm x 196.85 mm)
4. To enter the range for the control, select [Set Range]. The format for entering
the control range is "lll-hhh," where "lll" is the low end and "hhh" is the high end
of the control range.
5. To confirm the entry, press [Enter].
6. To return to the Quality Control Menu, select [Exit to Quality Control Menu].
To verify/record the ACT Plus™ actuator heat block temperature:
From the Quality Control Menu:
1. Select [Temperature Adjustment]. The following options will appear:
[Thermometer Reading], [Transmit Temperature Log], [Repeat
Adjustment], and [Exit to Quality Control Menu]. The Thermometer
Reading field is active when this screen is entered.
2. Enter the reading from the temperature verification thermometer using the
numeric keypad.
3. To confirm the entry, press [Enter].
4. To repeat the temperature adjustment (see Section 9 for Actuator Heat Block
Temperature), select [Repeat Adjustment].
5. To transmit the temperature log to the output location selected in the
Instrument Parameters, select [Transmit Temperature Log].

To view the due status for quality control testing:

From the Quality Control Menu:
1. Select [QC Due Status]. The ACT Plus™ instrument will display all liquid and
electronic controls for the cartridges that have been enabled with the following
message options:
The date and time when the next control tests are due.
QC due, if the controls for a test are past due.
QC not performed, if controls have never been run for a cartridge type.

QC Manager Functions

The QC Manager Setup Menu is accessed from the Quality Control Menu; it is a
password-protected area. The default password is provided in a separate
envelope. If this card is lost, contact Medtronic Technical Service for assistance
(800-328-3320).
The following variable function keys appear in the QC Manager Setup Menu:
[Location], [User ID Setup], [Clear ID Interval], [QC Lockout], [QC Interval
Setup], [Cartridge Enable Setup], [Download Settings], [Perm Record], and
[Instrument Upgrade].
Note: QC Manager parameters must be set before general QC information is
entered and before any tests are performed.
28 English
86702-001
Rev 1.0
Medtronic Confidential
ACT Plus™ Operator's Manual
CS010

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis