Message
Searching for reference position.
Protection hood not closed.
Emergency stop is active.
Out of Cutting Range.
Cutting finished.
Process stopping.
Error Message
Main supply voltage too low.
Cut-off motor overloaded.
ExciCut motor overloaded.
Protection hood not locked.
Cut-off motor blocked.
Contact K3 not activated.
Contact K4 not activated.
Step motor or encoder error.
Reference sensor not activated.
Step motor may be blocked.
AxioCut motor blocked.
Recirculation pump overloaded.
40
Exotom-100
Instruction Manual
Explanation
When unit is switched on, a
reference position process is
performed.
Wheel at bottom position when
Start is pressed.
Message after each cutting which
has been completed normally.
Protection hood can be opened.
Appears after Stop has been
pressed once.
Explanation
Supply voltage insufficient.
Wait until the motor has cooled
down.
Wait until the motor has cooled
down.
Safety lock has failed to lock hood.
Cut-off motor may fail to move the
wheel if caught in the workpiece.
Contact fails to operate.
Contact fails to operate.
Exotom-100 cannot complete
reference point search.
Search for reference point cannot
be conducted.
Cutting arm cannot move, most
likely due to blockage of arm.
Cutting arm cannot move, most
likely due to blockage of arm or
AxioCut motor may fail to move the
wheel if gets caught in the
workpiece.
Water hose blocked or too much
swarf in coolant.
Action
Wait until process has stopped.
Close hood properly.
Release emergency stop(s).
Move cut-off wheel to higher start
position.
Action
Await normalisation of power grid.
Press Start.
Press Start.
Re-close hood.
Move cut-off wheel clear of
workpiece before pressing Start.
Re-start.
If error remains, contact Struers
service technician.
Re-start machine using Start and
Stop buttons. Alternatively, re-start
using Main Switch
If error remains, contact Struers
service technician.
Re-start using Main Switch.
If error remains, contact Struers
service technician.
Check whether cutting arm is
blocked. Re-start using Main
Switch.
If error remains, contact Struers
service technician.
Remove blockage.
If error remains, contact Struers
service technician.
Move obstacle or move cut-off
wheel clear of workpiece before
pressing Start.
Replace coolant. Inspect pump.