WARNING!
Risk of electrical shock
To avoid the risk of electrical shock, only use this device when connected to a
properly grounded power supply network.
2.1.2
Precautions
CAUTION!
When using morcellation/shaver systems, the combination of low set pressures
and high vacuum pressures may result in a significant loss of distension pressure
which has the potential to affect the visibility of the surgical field. Conversely,
when employing high distension pressures, the deactivation of a morcellator
system can lead to pressure spikes that can exceed 150 mmHg.
CAUTION!
Mains voltage
Check to make sure the available mains voltage matches the data listed on the
type label attached to the back of the device. Incorrect voltage can cause errors
and malfunctions and may destroy the device.
CAUTION!
Indoor climate
Before switching on the device, sufficient time must have passed to adjust to the
indoor climate.
CAUTION!
Endoscope
The device may only be connected with endoscopes designed for and featuring
the technical specification permitting such a combined use. Any utilized endo-
scopes must comply with the most recent versions of IEC 60601-2-18 and ISO
8600. Combining/connecting with other devices generates a medical electrical
system (MES). The system configurator is responsible for compliance with the
standard IEC 60601-1 / EN 60601-1 in its latest version.
CAUTION!
Electrical Interference
(See Chapter 17 "Electromagnetic Compatibility"). Care was taken during the de-
velopment and testing of this device that electrical interference from this device
or other devices or instruments was practically eliminated. However, if you still
detect or suspect such interference, please follow these suggestions:
• Move this, the other or both devices to a different location
• Increase distance between used devices
• Consult an electro-medical expert