Command
Function
T
Tare
T-100 Set tare value to -100
I0
Integration time 0,08 sec.
I1
Integration time 0,16 sec.
I2
Integration time 0,32 sec.
S0
Stability low
S1
Stability middle
S2
Stability high
P0
Send single vaule
P1
Send single stable value
P2
Send single stable value automatically after load change
P3
Send value continuosly after every integration time
PDT
Print date and time (only if real clock installed)
R%
Set reference value to 100%
R1P
Set reference value to 1 Pieces
R10P Set reference value to 10 Pieces
R25P Set reference value to 25 Pieces
R100P Set reference value to 100 Pieces
R200P Set reference value to 200 Pieces
UC
Switch over from Unit1 to Unit2 and vice versa
CAL
Calibrate balance (only if INT or EXT selected
D
Write on display (see examples below)
DN
Display of weight values
N
Balance is no more remote-controlled
ON
To switch on the balance
OFF
To switch off the balance
PST
Statusprint
@
Alphanumeric display describes as example @ PRECISA BALANCE
@@
To switch the Alphanumeric display to an Analogue display
To alter the following menus please refer to the Operating handbook on page 8,9,10.
BAUDRATE:
PARITY
SYNCHRONIZATION:
PAG OERLIKON AG
SERIES-480
Baudrate = 300/600/1200/24004800/9600 Baud
To set the Baudrate in the configuration menu refer to page 8 operating
handbook (page 8)
Parity = odd, even and no
To set Parity in the configuration menu refer to operating handbook page 9
Print delay = 0,5 sec., 1,0 sec. and 2,0 sec. or Software-Handshake
Xon/Xoff
To set the Print delay in the configuration menu refer to operating
handbook page 10
SERIAL INTERFACE RS 232/V24 IP65 (350-8610)