E. Carefully slide the flexible, distal portion of the catheter out of the
hemostatic adapter, and close it onto the guidewire to hold it securely
in place. Completely remove the catheter from the guidewire while
maintaining guidewire position across the stenosis.
Table 2. Typical NC Emerge™ PTCA Dilatation Catheter Balloon Compliance
Pressure
atm
kPa
2.00
3.0
304
1.75
4.0
405
1.80
5.0
507
1.84
6.0
608
1.87
7.0
709
1.90
8.0
811
1.93
9.0
912
1.95
10.0
1013
1.97
11.0
1115
1.99
12.0
1216
Nominal
2.00
13.0
1317
2.01
14.0
1419
2.03
15.0
1520
2.04
16.0
1621
2.05
17.0
1723
2.06
18.0
1824
2.06
19.0
1925
2.07
20.0
2027
2.08*
*Rated Burst Pressure and Stent Rated Burst Pressure. DO NOT EXCEED.
**Rated Burst Pressure. DO NOT EXCEED.
REFERENCES
The physician should consult current medical text books on PTCA such as that
published by the American College of Cardiology/American Heart Association.

WARRANTY

Boston Scientific Corporation (BSC) warrants that reasonable care has been
used in the design and manufacture of this instrument. This warranty is in lieu
of and excludes all other warranties not expressly set forth herein, whether
express or implied by operation of law or otherwise, including, but not limited
to, any implied warranties of merchantability or fitness for a particular purpose.
Handling, storage, cleaning and sterilization of this instrument as well as other
factors relating to the patient, diagnosis, treatment, surgical procedures and
other matters beyond BSC's control directly affect the instrument and the results
Balloon size (mm)
2.25
2.50
2.75
3.00
1.95
2.19
2.38
2.65
2.00
2.25
2.44
2.71
2.04
2.29
2.48
2.76
2.08
2.34
2.53
2.81
2.12
2.38
2.57
2.86
2.15
2.41
2.61
2.90
2.17
2.44
2.64
2.93
2.19
2.47
2.67
2.96
2.21
2.49
2.69
2.99
2.23
2.51
2.71
3.01
2.25
2.52
2.73
3.03
2.26
2.54
2.75
3.04
2.27
2.55
2.76
3.06
2.28
2.56
2.78
3.08
2.29
2.58
2.79
3.09
2.30
2.59
2.80
3.10
2.31
2.60
2.82
3.12
2.33*
2.61*
2.83*
3.13*
F. Prepare the next catheter to be used as described in the Catheter
Preparation section.
G. Back load the new catheter onto the guidewire as described under step 2
Catheter Advancement and continue the procedure.
3.25
3.50
3.75
4.00
2.87
3.03
3.28
3.59
2.94
3.11
3.35
3.67
3.00
3.17
3.41
3.74
3.05
3.23
3.47
3.81
3.10
3.29
3.53
3.87
3.14
3.33
3.58
3.92
3.18
3.37
3.62
3.96
3.21
3.41
3.66
3.99
3.23
3.44
3.69
4.03
3.25
3.46
3.72
4.05
3.27
3.49
3.74
4.08
3.29
3.51
3.77
4.10
3.31
3.53
3.78
4.12
3.32
3.54
3.80
4.14
3.34
3.56
3.82
4.16
3.35
3.57
3.84
4.17
3.36
3.59
3.85
4.19
3.38*
3.61*
3.87*
4.21*
obtained from its use. BSC's obligation under this warranty is limited to the repair
or replacement of this instrument and BSC shall not be liable for any incidental
or consequential loss, damage or expense directly or indirectly arising from the
use of this instrument. BSC neither assumes, nor authorizes any other person to
assume for it, any other or additional liability or responsibility in connection with
this instrument. BSC assumes no liability with respect to instruments reused,
reprocessed or resterilized and makes no warranties, express or implied,
including but not limited to merchantability or fitness for a particular purpose,
with respect to such instruments.
7
4.50
5.00
5.50
6.00
4.02
4.49
4.86
5.40
4.12
4.59
4.96
5.51
4.20
4.68
5.05
5.62
4.28
4.76
5.14
5.72
4.34
4.84
5.22
5.81
4.40
4.90
5.29
5.89
4.45
4.95
5.35
5.95
4.49
5.00
5.40
6.00
4.52
5.03
5.44
6.05
4.55
5.07
5.48
6.09
4.58
5.10
5.52
6.13
4.61
5.13
5.55
6.17
4.63
5.15
5.58
6.20
4.65
5.17
5.60
6.22
4.67
5.19
5.62
6.25
4.70*
5.22*
5.65**
6.28**
Black (K) ∆E ≤5.0