PNEUMATIC CYLINDER REPLACEMENT

Tools Required:
• (2) 17 mm Combination Wrenches
• 11/16" Combination Wrench
• 5/16" Allen Wrench
Procedure:
1 .
Apply the stretcher brakes .
2 .
Raise the litter to its highest height .
3 .
Raise the fowler to 70° .
4 .
Using two 17 mm wrenches, remove the two nuts from the top of the pneumatic cylinder .
5 .
Using a 5/16" Allen wrench and an 11/16" combination wrench, remove the nut and bolt from the bottom of the
cylinder .
6 .
Raise the fowler to 90° and remove the cylinder .
7 .
Press down on the release pin on the replacement cylinder to extend the cylinder .
8 .
Reverse steps to install the new cylinder .
9 .
Check the adjustment of the pneumatic cylinders . See "Pneumatic Fowler Adjustment" on
10 . Verify proper operation before returning the unit to service .
www.stryker.com
Service Information
1089-009-006 REV G
page
1-29 .
Return To Table of Contents