Herunterladen Inhalt Inhalt Diese Seite drucken

Vent Device Installation; Vibration Isolator Installation - Varian Turbo-V300 ICE Series Bedienungshandbuch

Inhaltsverzeichnis

Werbung

VENT DEVICE INSTALLATION

The vent device mod. 969-9831 allows to avoid
undesired venting of the pump during a temporary
power failure (adjustable time up to 36 min.), and
enables an automatic vent operation.
Unscrew and remove the threaded plug (see figure
below).
Screw the flange mod. 969-9109 on the pump, tak-
ing care of the o-ring right position.
Assemble the seal ring and lock the vent device in
position using the KF klamp.

VIBRATION ISOLATOR INSTALLATION

Two vibration isolators for ISO flange are available
as accessories.
The two model part numbers are the following:
− model 969-9342 for ISO 100 flange;
− model 969-9343 for ISO 160 flange.
They typically reduce the vibration transmitted from
the Turbo-V300 ICE pump to the system by a fac-
tor of 20.
Please refer to the relevant instruction manual.
31
TECHNICAL INFORMATION
87-900-897-01(C)

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

969-9076969-9083969-9095969-9096

Inhaltsverzeichnis