BIOTRONIK Reocor S Gebrauchsanweisung Seite 87

Externer herzschrittmacher
Vorschau ausblenden Andere Handbücher für Reocor S:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
88
Test of
Test level
resistance to
according to
interference
IEC 60601
±6 kV contact dis-
Electrostatic
discharge (ESD)
charge
according to
±8 kV air discharge
IEC 61000-4-2
Fast transient
Not applicable
electric inter-
ference/bursts
according to
IEC 61000-4-4
Surges voltages 
Not applicable
(surges)
according to
IEC 61000-4-5
Voltage drops,
Not applicable
brief interrup-
tions and fluc-
tuations in the
supply voltage
according to
IEC 61000-4-11
Magnetic field
3 A/m
at the supply
frequencies
(50/60 Hz)
according to
IEC 61000-4-8
Conformity According to IEC 60601-1-2
Manufacturer guidelines and declaration –
resistance to electromagnetic interference 
(IEC 60601-1-2: Table 2)
is intended for use in an electromagnetic
The device
environment as described below. The user of the
device should make sure that it is used in such an
environment.
Compliance
level
±6 kV contact
discharge
±15 kV air
discharge
30 A/m
Guidelines for the electro-
magnetic environment
Floors should be made of
wood, cement or ceramic tiles.
When the floor consists of a
synthetic material, the relative
humidity must be at least 30%.
The magnetic field strength
should correspond to the
typical value in business and
hospital environments.