Agilent Technologies 4UHV Gebrauchsanleitung Seite 63

Ion pump controller

Werbung

Tab. 5
Pump type
500 diode
300 diode
150 diode
75-55-40 diode
20 diode
10 diode
500 StarCell
300 StarCell
150 StarCell
75-55-40 StarCell
20 StarCell
Set-point
The controller has 4 relays (relay A, relay B, relay C, relay D) and
they can be configured as set-points.
Set-point configuration
You can configure the four relays to operate with any channels.
Example 1 (default configuration): one relay for each channel.
Relay A  channel 1; Relay B  channel 2; Relay C  channel 3;
Relay D  channel 4.
Example 2: two relays for one channel.
Relay A  channel 1; Relay B  channel 1 ; Relay C  channel 3;
Relay D  channel 3.
Example 3: all relays for only one channel (e.g. channel 2).
Agilent 4UHV Ion Pump Controller User Manual
Technical Information
Protect current
100mA
100mA
50mA
30mA
20mA
5mA
100mA
100mA
50mA
30mA
20mA
/
87-900-137-01 (D)
5
Start-Up
63/114