Diagnostic Buffer (15 byte)
Tab. 25
Offset
Meaning
0
B
station non exist (set by master)
0
B
station not ready (slave not ready for data exchange)
1
B
cfg fault (configuration data doesn't match)
2
B
ext diag (slave has external diagnostic data)
3
B
not supported (slave doesn't support requested function)
4
B
invalid slave response ( slave sets permanent '0')
5
B
prm fault (wrong parameter assignment)
6
B
master_lock (slave is parameterised by another master)
7
1
B
prm req (slave has to be re-parameterised)
0
B
stat diag (static diagnosis)
1
B
fixed '1'
2
B
wd_on (response monitoring active)
3
B
Freeze mode (received freeze command)
4
B
Synch mode (received synch command)
5
B
reserved
6
B
diag deactivated (slave is parameterised by another master)
7
2
Reserved
3
Master Add (master address after parameterisation. FF= without parameter)
Ident number high
4
5
Ident number low
6
External diagnosis "Header" length indication including header
Agilent 4UHV Ion Pump Controller User Manual
Technical Information
/
87-900-137-01 (D)
5
PROFIBUS
105/114