Herunterladen Diese Seite drucken

Agilent Technologies 4UHV Gebrauchsanleitung Seite 41

Ion pump controller

Werbung

The 4UHV controller belongs to the second installation category (or overvoltage)
CAUTION!
as provided for in normative EN 61010-1. Therefore the device must be
connected to a power supply line satisfying the requirements for this category.
The 4UHV controller has connectors for inputs/outputs and for serial
communications that must be connected to the external circuits in such a way
that no powered part is accessible. Make sure that the insulation of the device
connected to the 4UHV controller provides an adequate insulation even in the
case of a single fault condition as stipulated in the EN 61010-1 normative.
If the controller is installed in a rack, remove its four feet and place it so as to
NOTE
leave at least 30 mm (1.2 inches) of free space above and below.
In rare circumstances, a controller fault could result in the emission of smoke. If
CAUTION!
the customer is using the controller in clean environments, they should take
adequate measures to avoid the environment being contaminated by any
eventual smoke.
The controller must be used in the following environmental
conditions:
Agilent 4UHV Ion Pump Controller User Manual
Temperature: 0 °C to +40 °C
Relative humidity: 0 – 90 % (non-condensing).
/
87-900-137-01 (D)
Instructions for Use
Installation
41/114
4

Werbung

loading